BIPOLAR
FORCEPS
Do you need a bipolar forcep for your electrosurgery
unit?
Macan replacement accessories will make
the difference.
|
|
|
|
BIPOLAR
FORCEPS (BPF-S1, BPF-C1, BPF-S2, BPF-C2):
Macan Bi-polar Forceps come in a variety of popular styles,
either straight or curved tips with or without serration.
Each Forceps comes with a special forceps cord to be used
with your Macan Bi-polar electrosurgery unit. The forceps
cord may also be purchased separately. |
|
|
Bipolar
Forceps
Why spend $6,000 on a
bi-polar machine?
|
|
Download
PDF
|
|
|
When you
choose Macan accessories you have chosen the best!
|
Cleaning
and care of Macan accessories
All accessories must be inspected for wear
and or damage every 30 days. This inspection will insure that
your accessories and the Macan electrosurgery unit are functioning
properly to meet the maximum performance and safety specifications.
If you find that an accessory part is worn or damaged, replace
it immediately. Accessories are not covered under the Macan
electrosurgery unit warranty.
All Macan accessories may be sterilized by standard methods.
All should be cleaned of blood and tissue debris before sterilization.
Standard methods of sterilization include:
Autoclave: 273 F @ 30# Pressure
Dry Heat: 320 F, 1 Hour
Chemical: Any bactericide
Handpieces and cords require regular sterilization. Autoclaving
is the preferred method. Handpieces should be inspected for
wear and damage of the plastic body of the handpiece, the
cable, and the plug that fits into the electrosurgery unit.
- The
handpiece should be inspected at the metal-to-plastic
joint under the handpiece cap. This joint is called the
seat. If any cracks are found in the plastic seat, the
handpiece should be replaced. This cannot be repaired.
- The
wire inside the handpiece shaft is prone to wear and breakage.
If this wire is broken inside the handpiece there will
be intermittent output to the electrode tips or no output
at all. To check this simply hold the handpiece body firmly
in one hand and pull sharply on the handpiece cord away
from the body of the handpiece with the other hand. If
the handpiece cord stretches or gives in any way, replace
the handpiece. This cannot be repaired.
- Care
should be taken to not lose the screw-on handpiece cap.
The handpiece will not hold the electrode tips in place
without this cap. If this screw-on cap is lost a replacement
can be purchased from Macan.
The Macan
ground plate and ground cord are both very important and delicate.
They provide a return current path back to the electrosurgery
unit. They must be in perfect working order at all times.
Inspect both the flat surface of the plate as well as the
edges of the plate for "nicks or breaks" in the insulation
material. If there is exposed metal on the plate, it should
not be used and replaced immediately. Inspect the ground cord
for cuts and breaks in the cable. The plugs on both ends should
be checked for wear and to make sure they are connected properly
to the cable. Do not use a defective gound plate or ground
cord with the electrosurgery unit.
Electrodes are sterile at the moment of contact with the tissue
because the high frequencies generated by the electrosurgery
unit destroy any organism within range. This does not
apply to the plastic sleeve that protects the shaft of the
electrode. The electrodes can be sterilized using any of the
standard methods.
|
|
|